Figure 13. Trace discontinuity at eighth graticule.
Sampler Bias Adjustment

Calibration Procedure For Digitizing Oscilloscope Hewlett Packard, Model 54121T, Model 54120B, And Model 54121A Page Navigation
  28    29    30    31    32  33  34    35    36    37    38  
img
. . . .
TB 9-4931-538-40
Figure 14. Expanded trace discontinuity.
Figure 15. Correct and adjustment - no discontinuity.
22. Sampler Bias Adjustment
a. Set the display menu and mainframe settings to meet the following conditions:
(1) Press Display key. Display Mode to Average is highlighted. NUMBER OF
AVERAGES to # Averages = 4. Screen to Single is highlighted. Graticule to Grid is
highlighted. Bandwidth to 12.4 GHz is highlighted.
33