Table 16. Amplitude Modulation Internal.
Amplitude Modulation External

Calibration Procedure For Synthesized Signal Generator Model Mg3695B, With Options 1B, 2B, 5, 6, And 28B Page Navigation
  16    17    18    19    20  21  22    23    24  
TB 9-6625-106-24
(4) At the prompt on the instrument controller screen, type:
calterm 112 <ENTER>.
(5) Set function generator output to 0.00 V dc, then press any key on the instrument
controller to continue calibration.
(6) Toggle the function generator output between +0.50 V dc and 0.50 V dc as
indicated on the multimeter.
(7) On the instrument controller use 1, 2, or 3 to increment and 8, 9, and 0 to
decrement the value of the power meters setting to obtain a 9.54 dB difference in reading
when the function generator output is toggled from +0.50 V dc and 0.50 V dc.
(8) On the instrument controller press "q" to quit.
(9) Repeat technique of (5) through (8) above at 5 GHz.
(10) If TI firmware is 3.36 or higher, perform (a) through (e) below:
(a) At the prompt on the instrument controller screen, type: calterm 172
<ENTER>.
(b) Set the function generator to output 0.00 V dc, press any key on the
instrument controller to continue calibration.
(c) Toggle the function generator output between +0.50 V dc and 0.50 V dc as
indicated on the multimeter.
(d) On the instrument controller use 1, 2, or 3 to increment and 8, 9, and 0 to
decrement the value of the power meters setting to obtain a 9.54 dB difference in reading
when the function generator output is toggled from +0.50 V dc and 0.50 V dc.
(e) On the instrument controller press "q" to quit.
(11) At the prompt on the instrument controller screen, type:
calterm 113 <ENTER>.
(12) Set measuring receiver to measure RF power at 1 GHz.
(13) Toggle the function generator output between +0.20 V dc and 0.20 V dc as
indicated on the multimeter.
(14) On the instrument controller use 1, 2, or 3 to increment and 8, 9, and 0 to
decrement the value of the power meters setting to obtain a 10.00 dB difference in reading
when the function generator's output is toggled.
(15) On the instrument controller press "q" to quit.
(16) Repeat technique of (12) through (15) above at 5 GHz.
(17) If TI firmware is 3.36 or higher perform (a) through (g) below:
(a) At the prompt on the instrument controller screen, type:
calterm 173 <ENTER>.
(b) Set the function generator to output 0.00 V dc, then press any key on the
instrument controller to continue calibration.
(c) Toggle the function generator output between +0.20 V dc and 0.20 V dc as
indicated on the multimeter.