• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Narrow )M Mode On
Table 16. Amplitude Modulation Internal.

Calibration Procedure For Synthesized Signal Generator Model Mg3695B, With Options 1B, 2B, 5, 6, And 28B Page Navigation
  14    15    16    17    18  19  20    21    22    23    24  
img
. . . . . . .
TB 9-6625-106-24
b. Adjustments. Perform paragraph 11 (b) Locked FM Mode On. After
adjustment repeat 11 a through 14 a.
15. Wide )M Mode On
a. Performance Check
(1) Press TI keys as listed in (a) through (e) below:
(a) System, Reset.
(b) Edit L1, 0, dBm.
(c) Edit F1, 5, GHz.
(d) Modulation, )M, Internal/External, select external.
(e) Press, on/off, turn on External )M.
(2) Allow measuring receiver to tune to TI carrier frequency, then measure PM
Deviation with PEAK + -/2 detector, no filters, and record PM Deviation reading as (N).
(3) On the TI press, on/off, turn off External )M.
(4) On the TI press, Internal/External, select Internal )M.
(5) On the TI press, Edit Deviation, 3, Enter.
(6) On the TI press, Mode, Wide, <Previous.
(7) On the TI press, on/off, turn on Internal )M.
(8) Set measuring receiver to measure PM Deviation with PEAK + detector, and
record FM Deviation reading as (S+N).
(9) Calculate actual PM deviation by using the following formula, Sactual = (S+N)-N.
(10) If the calculated PM deviation is not within limits listed in table 15, perform b
below.
(11) On TI turn off Internal PM, then repeat (1) (c) through (e) and (3) through (10)
above for the remaining frequency listed in table 15. If the calculated PM deviation is not
within limits listed in table 15, perform b below.
(12) Set TI outputs to minimum.
Table 15. Wide )M Mode On.
PM Deviation
TI Carrier Frequency
(GHz)
Min (rad)
Max (rad)
5
2.70
3.30
20
2.70
3.30
b. Adjustments. Perform paragraph 11 (b) Locked FM Mode On. After adjustment
repeat 11 a through 15 a.
20


Privacy Statement