Unlocked Narrow FM Mode On
Wide )M Mode On

Calibration Procedure For Synthesized Signal Generator Model Mg3695B, With Options 1B, 2B, 5, 6, And 28B Page Navigation
  13    14    15    16    17  18  19    20    21    22    23  
img
. . . . . . . . .
TB 9-6625-106-24
Table 13. Unlocked Narrow FM Mode On
Frequency Deviation
TI Carrier Frequency
(GHz)
Min (kHz)
Max (kHz)
5
270
330
20
270
330
b. Adjustments. Perform paragraph 11 (b) Locked FM Mode On. After adjustment
repeat 11 a through 13 a.
14. Narrow )M Mode On
a. Performance Check
(1) Press TI keys as listed in (a) through (e) below:
(a) System, Reset.
(b) Edit L1, 0, dBm.
(c) Edit F1, 5, GHz.
(d) Modulation, )M, Internal/External, select external.
(e) Press, on/off, turn on External )M.
(2) Allow measuring receiver to tune to TI carrier frequency, then measure PM
Deviation with PEAK + -/2 detector, no filters, and record PM Deviation reading as (N).
(3) On the TI press, on/off, turn off External )M.
(4) On the TI press, Internal/External, select Internal )M.
(5) On the TI press, Edit Deviation, 3, Enter.
(6) On the TI press, Mode, Narrow, <Previous.
(7) On the TI press, on/off, turn on Internal )M.
(8) Set measuring receiver to measure PM Deviation with PEAK + detector, and
record PM Deviation reading as (S+N).
(9) Calculate actual PM deviation by using the following formula, Sactual = (S+N)-N.
(10) If the calculated PM deviation is not within limits listed in table 14, perform b
below.
(11) On TI turn off Internal PM, then repeat (1) (c) through (e) and (3) through (10)
above for the remaining frequency listed in table 14. If the calculated PM deviation is not
within limits listed in table 14, perform b below.
(12) Set TI outputs to minimum.
Table 14. Narrow )M Mode On
PM Deviation
TI Carrier Frequency
(GHz)
Min (rad)
Max (rad)
5
2.70
3.30
20
2.70
3.30