TB 9-6625-1098-24
NOTE
It may be necessary to reinstall the shorting bar between TI
and chassis ground terminals below 1 kHz.
(f) MODE switch to MANUAL (Hewlett-Packard Models 333A and 334A).
(3) Adjust calibrator frequency for 5 kHz and amplitude for an indication of 3 on TI
meter 0 to 3 scale.
(4) Set FUNCTION switch to DISTORTION.
(5) Adjust BALANCE controls and calibrator frequency for a null indication on TI
meter (METER RANGE switch may be reduced to obtain best null indication). If
calibrator frequency does not indicate between 4.75 and 5.25 kHz, perform b below.
(6) Set METER RANGE switch to .3 VOLTS and FUNCTION switch to
SET LEVEL.
(7) Repeat (2) through (6) above, using frequencies, control settings, and indications
Test instrument
Calibrator frequency
Nominal
FREQUENCY
indications
FREQUENCY
frequency
RANGE
dial settings
switch settings
Min
Max
20
kHz
X1K
20
19.0
kHz
21.0 kHz
60
kHz
X1K
60
57.0
kHz
63.0 kHz
50
kHz
X10K
5
47.5
kHz
52.5 kHz
200
kHz
X10K
20
190
kHz
210 kHz
600
kHz
X10K
60
540
kHz
660 kHz
6
kHz
X100
60
5700
Hz
6300 Hz
2
kHz
X100
20
1900
Hz
2100 Hz
500
Hz
X100
5
475
Hz
525 Hz
50
Hz
X10
5
47.5
Hz
52.5 Hz
200
Hz
X10
20
190
Hz
210 Hz
600
Hz
X10
60
570
Hz
630 Hz
60
Hz
X1
60
57
Hz
63 Hz
20
Hz
X1
20
19
Hz
21 Hz
b. Adjustments
(1) Position switches as listed in (a) through (c) below:
(a) METER RANGE to .3 VOLTS.
(b) FUNCTION to SET LEVEL.
(c) FREQUENCY RANGE to X1K.
(2) Adjust calibrator frequency for 5 kHz.
(3) Set FUNCTION switch to DISTORTION.
12