TB 9-6625-1098-24
Test instrument
Test instrument
Calibrator
FREQUENCY
FREQUENCY
(dB)
frequencies
RANGE
dial
switch settings
settings
Min
Max
X10K
20
200 kHz
-2.0
0.0
X10K
5
50 kHz
-1.0
0.0
X1K
5
5 kHz
-0.6
+0.6
X1K
15
15 kHz
-0.6
+0.6
X100
20
2 kHz
-0.6
+0.6
X100
5
500 Hz
-0.6
+0.6
X10
5
50 Hz
-0.6
+0.6
X10
20
200 Hz
-0.6
+0.6
X1
15
15 Hz
0.0
1.0
b. Adjustments. No adjustments can be made.
16. Automatic Null (Hewlett-Packard, Models 333A and 334A)
a. Performance Check
(1) Connect calibrator OUTPUT to TI INPUT.
(2) Position TI controls as listed in (a) through (g) below:
(a)
METER RANGE switch to 1 VOLT.
(b)
MODE selector switch to MANUAL.
(c)
FREQUENCY RANGE switch to X100.
(d)
(e)
SENSITIVITY switch to MAX.
(f)
SENSITIVITY VERNIER control to MIN.
(g)
(3) Adjust calibrator for 5 kHz and adjust level for a 1 V indication on TI meter.
(4) Set FUNCTION and METER RANGE switches to SET LEVEL.
(5) Adjust SENSITIVITY and VERNIER controls for a +2 dB indication on
TI meter.
(6) Set FUNCTION switch to DISTORTION.
(7) Adjust FREQUENCY dial and BALANCE controls for a null indication on
TI meter.
(8) Reduce METER RANGE switch as necessary to maintain an on-scale reading.
Record TI meter reading.
(9) Set METER RANGE switch to 100% (SET LEVEL).
(10) Adjust FREQUENCY dial and BALANCE controls for a null indication on TI
meter. When TI indication is less than 3% on 0 to 3% (SET LEVEL) scale, set MODE
switch to AUTOMATIC.
(11) Reduce METER RANGE switch as necessary to maintain an on-scale reading.
Record TI meter reading.