TB 9-6625-163-24
Table 1. Calibration Description Continued.
Test instrument
Performance specifications1, 2
parameters
Range: 0.5 Hz to 999.99 kHz, Pulse width > 0.5 S
Range
Resolution
Accuracy
99.999 Hz
0.001 Hz
0.02% + 5
999.99 Hz
0.01 Hz
9.9999 kHz
0.0001 kHz
0.005% + 5
99.999 kHz
0.001 kHz
999.99 kHz
0.01 kHz
0.01 %
0.2% per kHz + 0.1%
Duty Cycle1,2,3
1.00% to 99.00%
Pulse Width1,2
0.1000 mS
0.0001 mS
0.002 mS + 30 counts
1.000 mS
0.001 mS
10.00 mS
0.01 mS
0.002 mS + 3 counts
1999.9 mS
0.1 mS
For rise times <1 μS. Signals centered around trigger levels.
1
0.5 to 200 kHz, pulse width >2 μs. Pulse width range is determined by the frequency of the signal.
2
3 For 10 μS < pulse width <25 μS add 1%. For 2 μS < pulse width ≤10 μS add 3.5 %.
Accuracy1
Temperature
Range
Readout
-200 C to +1350 C
(1% + 1 C)
(-328F to +2462 F)
(1% + 1.8 F)
Accuracy specification does not include the error of the thermocouple probe.
1
Accuracy1
Temperature
Range
(80PK-26 Probe)
-40 C to 293 C, (-40 F to 559.4 F)
2.2 C, (3.96F)
293 C to 816 C, (559.4 F to 1500 F)
0.75 %
overall accuracy of the combination using √ + . This was applied to the minimum use
TI Accuracy must be considered along with the 80PK-26 accuracy specification to determine the
1
specification in table 2 and applicable test limits.
Accuracy1
High Voltage
Range
(80K-6 Probe)
0-6 kV dc or Peak ac
DC to 500 Hz
500 Hz to 1 kHz
1%
2%
Above 1 kHz output reading falls. Typically, -30% at 10 kHz.
1
Accuracy specifications are given as: ( [ % of reading ] + [ number of least significant digits ] ).
1
2 Accuracy assumes ambient temperature stable to 1.8F (1C). For ambient temperature changes of 9F (5C),
rated accuracy applies after 2 hours.
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286; AN/GSM-287; or AN/GSM-440. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parentheses.