Table 3. Dc Voltage Accuracy.
Ac Current

Calibration Procedure For Multimeter Set, An/Gsm-437 And Digital Multimeter Model 287 Page Navigation
  3    4    5    6    7  8  9    10    11    12    13  
img
. . . . . . . . . . . . .
TB 9-6625-163-24
9. Ac Voltage
a. Performance Check
(1) Ensure that calibrator OUTPUT HI, LO is connected to TI V Ω and COM inputs.
(2) Set TI function switch to mV~.
(3) Set calibrator for an output of 5 mV at a frequency of 20 Hz, OPR. If TI does not
indicate within limits specified in first row of table 4, perform b below.
(4) Repeat technique of step (3) above for settings and indications listed in table 4.
If TI does not indicate within limits specified in table 4, perform b below.
Table 4. Ac Voltage Accuracy.
Calibrator output
Test instrument
Amplitude
Frequency
Range/Function
Min
Max
5 mV
20 Hz
50  mV
4.865  mV
5.135  mV
50 mV
65 kHz
48.210  mV
51.790  mV
100 kHz
500  mV
47.85
mV
52.15
mV
250 mV
65 kHz
240.85
mV
259.15
mV
500 mV
45 Hz
498.25
mV
501.75
mV
1
100 mV
60 Hz
5V
0.0952 V
0.1048  V
500 mV
10 kHz
0.4945 V
0.5055  V
3V
100 kHz
2.8160 V
3.1840  V
15 V
50  V
14.435  V
15.565
V
500 V
10 kHz
500  V
497.75
V
502.25
V
1000 V
1000  V
993.5
V
1006.5
V
Set TI function switch to V~.
1
(5) Set calibrator output to minimum and disconnect equipment.
b. Adjustments. Complete all performance checks in paragraphs 8 through 14,
recording failures, then perform entire alignment procedure in paragraph 17.
10. Dc Current
a. Performance Check
(1) Connect calibrator OUTPUT HI, LO to TI mA/A and COM inputs.
(2) Set function switch to A~
.
(3) Set calibrator output for 500 A, OPR. If TI does not indicate within limits
specified in first row of table 5, perform b below.
(4) Repeat technique of step (3) above, using settings and indications listed in
table 5. If TI does not indicate within limits specified in table 5, perform b below.
Table 5. Dc Current Accuracy.
Calibrator
Test instrument
Output
Range/Function
Min
Max
500 A
500  A
499.42
A
500.58
A
5000 A
5000 A
4996.0
A
5004.0
A
1
0.1  mA
50
mA
0.090  mA
0.110  mA
50  mA
49.965  mA
50.035  mA
400 mA
400  mA
399.38
mA
400.62
mA
A2
5
5
A
4.9840 A
5.0160 A
10  A
10
A
9.968  A
10.032  A
Set function switch to mA/A~
.
1
Set calibrator to standby. Connect calibrator amplifier CURRENT OUTPUT HI, LO to TI A and COM inputs.
2
9