TB 9-6625-163-24
Table 8. Calibration Adjustment Steps For Firmware Version 1.10.
Calibrator
Test instrument
Rotary switch
Settling time
Step #
time
Output
position
(Seconds)
(Seconds)
1
2
7
0.0 V
0 Hz
mV DC
2
3
6
500.00 mV
0 Hz
mV DC
3
2
8
0.0 mV
0 Hz
mV DC
4
2
1
50.0 mV
0 Hz
mV DC
5
2
1
100.00 mV
0 Hz
mV DC
6
2
1
500.00 mV
0 Hz
mV DC
7
4
2
5.0000 mV
700 Hz
mV AC
8
3
3
50.000 mV
700 Hz
mV AC
9
3
2
500.00 mV
700 Hz
mV AC
10
2
2
500.00 mV
55Hz
mV AC
11
10
6
50.000 V
55 Hz
mV AC
0.0000 V
0 Hz
mV AC
12
5
35
13
2
15
0.000 V
0 Hz
V AC
14
3
7
5.0000 V
55 Hz
V AC
15
3
42
5.0000 V
60 kHz
V AC
16
2
10
50.000 V
55 Hz
V AC
17
4
94
50.000 V
100 kHz
V AC
18
5
2
500.00 V
55 Hz
V AC
19
3
2
1000.0 V
55 Hz
V AC
20
2
8
0.0000 V
0 Hz
V DC
21
2
1
5.0000 V
0 Hz
V DC
22
2
1
50.000 V
0 Hz
V DC
23
2
1
500.00 V
0 Hz
V DC
24
3
1
1000.0 V
0 Hz
V DC
25
3
12
5.0000 V
55 Hz
V DC
26
3
13
5.0000 V
20 kHz
V DC
27
5
16
50.000 V
55 Hz
V DC
28
5
32
50.000 V
20 kHz
V DC
29
5
7
1000.0 V
55 Hz
V DC
30
5
8
500.00 mV
55 Hz
mV DC
0 Ω1
Ω
31
2
15
500 Ω2
Ω
32
3
1
5 kΩ
Ω
33
3
1
50 kΩ
Ω
34
2
1
500 kΩ
Ω
35
2
13
5.0 MΩ3
Ω
36
2
9
50.0 MΩ
Ω
37
2
9
500.0 MΩ
Ω
38
2
2
0 Ω1
39
2
1
3.1 V4
40
2
1
0 Hz
5000.0 A5
41
2
1
0 Hz
A DC
42
2
1
5000.0 A
55 Hz
A AC
43
2
1
400.0 mA
0 Hz
mA AC
44
2
1
400.0 mA
55 Hz
mA AC
10.0 A6
45
2
1
0 Hz
A DC
46
2
2
10.0A
55 Hz
A AC
Connect high quality short banana lead to TI V Ω, COM input.
1
Connect resistance standard 1 CURRENT HI, LO to TI V Ω, COM input.
2
3 Connect resistance standard 2 H, L to TI V Ω, COM input.
4 Connect calibrator OUTPUT HI, LO to TI V Ω, COM input.
5 Connect calibrator OUTPUT HI, LO to TI mA/A and COM input.
6 Disconnect banana lead from TI mA/A and connect to A.
15