TB 9-6625-166-24
WARNING
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Set TI rotary switch to DCV 2.
b. Plug the red test lead into the TEMP V R terminal and the black test lead into
COM terminal.
c. Ensure calibrator is set to standby.
a. Performance Check
(1) Connect the TI TEMP V R and COM terminals to the calibrator.
(2) Set calibrator output for the first value in table 3. TI will indicate within limits
specified in first row of table 3. If TI does not indicate within limits specified in table 3,
perform b below.
(3) Repeat technique of (2) above, for the remaining settings and indications listed
in table 3. TI will indicate within limits specified in table 3. If TI does not indicate within
limits specified in table 3, perform b below.
Table 3. Dc Voltage Accuracy.
RANGE (DCV)
Calibrator Output
Min (V)
Max (V)
2
1.9 V
1.876
1.924
20
5V
4.93
5.07
20
10 V
9.87
10.13
20
-10 V
-9.87
-10.13
20
15 V
14.81
15.19
20
19 V
18.76
19.24
200
190 V
187.6
192.4
600
590 V
582
598
(4) Set calibrator output to minimum.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Set TI rotary switch to ACV 200V.
(2) Set calibrator output for the first value in table 4. TI will indicate within limits
specified in first row of table 4. If TI does not indicate within limits specified in table 4,
perform b below.
(3) Repeat technique of (2) above for the remaining settings and indications listed in
table 4. If TI does not indicate within limits specified in table 4, perform b below.