TB 9-6625-1947-35
-R15-
Figure 1. Adjustment locations -
bottom view.
(2) Connect audio analyzer or spectrum analyzer to TI UNBAL output, using 50 Ω
feed-through termination.
(3) Position controls as listed in (a) through (d) below:
(b) FREQUENCY dial fully cw.
(c) OUTPUT LEVEL dBm switch to +10.
(d) AMPLITUDE control fully ccw.
(4) Adjust A2R3 (fig. 1) for a -350 mV indication on multimeter (R).
(5) Adjust A2R4 (fig. 1) for minimum distortion on audio analyzer, or at least -46 dB
on the spectrum analyzer (R).
(7) Alternately adjust S2C2 and S2C8 (fig. 2) until frequency counter indicates 10.2
kHz and multimeter indicates between -340 and -360 mV (R).
6