TB 9-6625-1947-35
(13) Adjust calibrator to -0.5 percent of value recorded in (11) above. Record
(14) Disconnect calibrator from thermal converter and connect thermal converter to
TI UNBAL output. Set FREQUENCY RANGE switch to X1M and FREQUENCY dial to 10.
(15) Adjust AMPLITUDE control for a 0 dBm indication on TI meter. If multimeter
does not indicate within limits of R3 and R4, perform b (11) through (13) below.
(16) Repeat technique of (14) and (15) above for FREQUENCY RANGE switch and
FREQUENCY dial
settings listed in
table 5. If
multimeter does not
indicate within
limits
of R3 and R4, perform b (11) through (13) below.
Test instrument
FREQUENCY RANGE
FREQUENCY
switch settings
dial settings
X1M
4
X1M
1
X100K
5
X10K
5
X1K
5
X100
1
X10
2
X10
1
b. Adjustments
(1) Rotate A3R18 (fig. 2) fully cw.
(2) Adjust A3R7 (fig. 2) until AMPLITUDE control has sufficient range to exceed
maximum meter deflection at both ends of meter scale.
(3) Adjust AMPLITUDE control for a -1 dBm indication on TI meter. Record
(4) Multiply R2 by 1.259 and record as R4.
(5) Adjust AMPLITUDE control for value of R4 on multimeter.
(6) Adjust A3R18 (fig. 2) for a +1.0 dBm indication on TI meter
(7) Adjust AMPLITUDE control for a 0.7071 V rms indication on multimeter.
(8) Adjust A3R7 (fig. 2) for a 0 dBm indication on TI meter (R).
(9) Rotate AMPLITUDE control fully cw.
(10) Adjust A3R20 (fig. 2) for a 0.80 V indication on multimeter (R).
(11) Adjust AMPLITUDE control for a 0 dBm indication on TI meter.
(12) Adjust A2C43 (fig. 1) for R2 value recorded in 9 a (9) above (R).
(13) Repeat 9 a (14) above.