TB 9-6625-1947-35
(10) Repeat technique of (5) through (8) above with resistance standard set to 300 Ω
between 2.424 and 2.474 V rms.
b. Adjustments. No further adjustments can be made.
a. Performance Check
(1) Connect audio analyzer or spectrum analyzer to TI UNBAL output connector,
using a 50 Ω feed-through termination.
(2) Position controls as listed in (a) through (d) below:
(a) FREQUENCY RANGE switch to X10.
(b) FREQUENCY dial to 1.
(c) OUTPUT LEVEL dBm switch to +10.
(3) Adjust AMPLITUDE control for a +1 dBm indication on TI meter. If measured
distortion is not greater than 40 dB below the fundamental frequency, perform b below.
(4) Repeat technique of (2) and (3) above at FREQUENCY RANGE switch and
FREQUENCY dial
settings listed in
table 6. Measured distortion will be
greater than 40
dB below fundamental frequency.
Test instrument
FREQUENCY RANGE
FREQUENCY
switch settings
dial settings
X10
5
X10
10
X100
1
X100
10
X1K
10
X10K
1
X10K
2
X100K
1
X100K
5
X100K1
10
X1M
10
X1M
5
X1M
1
1Replace
audio analyzer with spectrum analyzer and measured distortion
will be greater than 34 dB below fundamental frequency.
b. Adjustments.
Adjust A2R4 (fig. 1)
for minimum
distortion of at
least 46 dB
(10 Hz
to 500 kHz) and at least -34 dB (1 to 10 MHz) below fundamental FREQUENCY (R).