TB 9-6625-1947-35
a. Performance Check
TI
BAL
UNBAL
RESISTANCE STANDARD
RESISTANCE STANDARD
MULTIMETER
Figure 3. Output impedance -
equipment setup.
(2) Position controls as listed in (a) through (d) below:
(c) OUTPUT LEVEL dBm switch to +10.
(3) Set both resistance standards to 75 Ω.
(4) Adjust AMPLITUDE control for a 0 dBm indication on TI meter. Multimeter
will indicate between .8573 and .8747 V rms.
(5) Set both resistance standards to 67.5 Ω.
(6) Press IMPEDANCE 135 BAL pushbutton.
(7) Adjust AMPLITUDE control for a 0 dBm indication on TI meter.
Sum of multimeter readings will be between 1.150 and 1.174 V rms.
(9) Repeat technique of (5) through (8) above with resistance standard set to 75 Ω
and press IMPEDANCE 150 BAL pushbutton. Sum of multimeter readings will be
between 1.213 and 1.237 V rms.