TB 9-6625-1947-35
12. Output Attenuator Accuracy
a. Performance Check
(1) Connect true rms voltmeter to UNBAL output connector, using a 50 Ω feed-
through termination.
(2) Position controls as listed in (a) through (e) below:
(c) OUTPUT LEVEL switch to +10 dBm.
(e) AMPLITUDE control to +0 dBm.
(3) True rms voltmeter will indicate between +9.85 to +10.15 dBm.
(4) Repeat technique of (2) and (3) above for settings and indications listed in table 7.
True rms voltmeter will indicate within specified limits.
Test instrument
True rms voltmeter indications
OUTPUT LEVEL
(dBm)
switch settings (dBm)
Min
Max
0
-0.15
+0.15
-10
-9.85
-10.15
-20
-19.85
-20.15
-30
-29.85
-30.15
-40
-39.85
-40.15
-50
-49.85
-50.15
-60
-59.85
-60.15
-1
-0.85
-1.15
-2
-1.85
-2.15
-3
-2.85
-3.15
-4
-3.85
-4.15
-5
-4.85
-5.15
-6
-5.85
-6.15
-7
-6.85
-7.15
-8
-7.85
-8.15
-9
-8.85
-9.15
b. Adjustments. No adjustments can be made.
a. Performance Check
NOTE
Do not perform power supply check if all other parameters are
within tolerance.