(25) Using TI cursor knob, set v2 cursor on deflected baseline. If TI ĆV indication at
bottom of TI display is not within limits specified in first row of table 4, perform b below.
(26) Set oscilloscope calibrator output to OFF.
(27) Repeat technique of (17) through (26) above for TI ranges and oscilloscope
calibrator outputs listed in table 4. If TI ĆV indications at bottom of TI display are not
within limits specified in table 4, perform b below.
Oscilloscope calibrator
Test instrument
Output
Volts/Div
Indication limits (V)
(V dc)
Min
Max
35.0
5
34.31
35.69
14.00
2
13.73
14.27
7.00
1
6.863
7.137
3.50
500 m
3.431
3.569
1.400
200 m
1.373
1.427
700
m
100 m
686.3 m
713.7 m
350
m
50 m
343.1 m
356.9 m
140.0 m
20 m
137.3 m
142.7 m
70.0 m
10 m
68.63 m
71.37 m
1
35.0 m
5m
34.15 m
35.85 m
1
14
m
2m
13.47 m
14.53 m
FS is 80 mV for TI 2mV and 5mV ranges.
1
NOTE
For models 54601A and 54602A continue with (30) below. For
model 54600A proceed to (57) below.
(28) Set oscilloscope calibrator output to OFF and disconnect oscilloscope calibrator
from TI channel 2 input and connect to TI channel 3 input.
(29) Press TI channel 2, as required to remove channel 2 from display. Press TI
channel 3.
(30) Set TI channel 3 to 500 mV range by pressing [.5].
(31) Set oscilloscope calibrator for a 3.5 V dc output with ground condition to ON.
(32) Adjust TI channel 3 position knob to place baseline near but not at bottom of display.
(33) Press TI Cursors, then press [v1].
(34) Using TI cursor knob, set v1 cursor on baseline.
(35) Set oscilloscope calibrator output to OFF.
(36) Set oscilloscope calibrator ground condition to off, then set oscilloscope calibrator
output to ON.
(37) Press TI [v2].
(38) Using TI cursor knob, set v2 cursor on deflected baseline. If TI ĆV indication at
bottom of TI display is not within limits specified in first row of table 5, perform b below.
(39) Set oscilloscope calibrator output to OFF.
7