TB 9-6625-1995-35
(23) Repeat technique of (18) through (22) above for settings listed in table 4. If
DELAY TIME MULT dial indication is not between 8.84 and 8.96 at each setting, perform
b below.
Test instrument
TIME/DIV OR DLY
DLY'D time/division
TIME
PUSH FOR DLY'D
marker
switch settings
switch settings
output settings
.1
s
2
ms
50
mS/D
50
ms
1
ms
10
mS/D
20
ms
.5
ms
10
mS/D
10
ms
.2
ms
5
mS/D
5
ms
.1
ms
1
mS/D
2
ms
s
1
mS/D
50
1
ms
s
.5
mS/D
20
s
.5
ms
.1
mS/D
10
.2
ms
s
.1
mS/D
5
.1
ms
s
S/D
2
50
s
s
S/D
50
1
10
s
s
S/D
20
.5
10
s
s
S/D
10
.2
5
s
s
S/D
5
.1
1
50
ns
s
S/D
2
1
s
20
ns
S/D
1
.5
10
ns
s
S/D
.5
.1
s
5
ns
50
nS/D
.2
5
ns
50
nS/D
s
.1
(24) Position controls as listed in (a) through (c) below:
(a) TIME/DIV OR DLY TIME switch to 50 ns.
(b) DLY'D time/division switch to 10 ns PUSH FOR DLY'D.
(c) MAIN TRIGGERING LEVEL control for stable display.
(25) Position vertical amplifier controls for approximately 2 divisions of display.
(26) Adjust DELAY TIME MULT dial to 9.00.
(27) Adjust POSITION control to aline the leading edge of time marker to the center
vertical graticule line.
(28) Decrease DELAY TIME MULT dial (near 8.0) to aline the leading edge of time
marker to the center vertical graticule line. DELAY TIME MULT dial will indicate
between 7.94 and 8.06.
(29) Repeat technique of (28) above for DELAY TIME MULT dial settings and
CHANGE 1 13