TB 9-6625-1996-24
a. Performance Check
(1) Set function generator as listed in (a) through (d) below:
Function Outputs to 50Ω.
(a)
Function to sine wave.
(b)
(c)
Amplitude to 0.001 VPP.
(d)
(2) Position TI controls as listed in (a) through (f) below:
CHANNEL A LEVEL control to PRESET.
(a)
CHANNEL A SLOPE switch to +.
(b)
CHANNEL A 50Ω/1MΩ switch to 50Ω.
(c)
CHANNEL A ATTEN X1/X10 switch to X1.
(d)
CHANNEL A AC/DC switch to DC.
(e)
CHECK/COM A/SEP switch to SEP.
(f)
(3) Connect TI CHANNEL A input to function generator Function Output
Unbalanced.
(4) Press the function generator Enter key to enable the knob, and slowly rotate the
function generator knob cw until TI display GATE light flashes and indication is stable at
approximately 10 Hz.
(5) Verify that the function generator indicates within limits specified in table 3. If
not perform b below.
(6) Repeat technique of (1) through (5) above for remaining settings listed in table 3.
Function generator/signal
Function generator/signal
CHECK-
Test
generator indications
COM A-SEP
instrument
(≤)
setting
switch setting
input
CHANNEL A
SEP
10 Hz
0.071 VPP
CHANNEL A
SEP
10 kHz
0.071 VPP
CHANNEL A
SEP
100 kHz
0.071 VPP
CHANNEL A
SEP
1 MHz
0.071 VPP
CHANNEL A1
SEP
10 MHz
0.025 VRMS
CHANNEL A
SEP
100 MHz
0.025 VRMS
CHANNEL A
SEP
500 MHz
0.025 VRMS
CHANNEL A2
COM A
10 Hz
0.141 VPP
CHANNEL A
COM A
10 kHz
0.141 VPP
CHANNEL A
COM A
100 kHz
0.141 VPP
CHANNEL A
COM A
1 MHz
0.141 VPP
CHANNEL A1
COM A
10 MHz
0.050 VRMS
CHANNEL A
COM A
100 MHz
0.050 VRMS
CHANNEL A
COM A
400 MHz
0.050 VRMS
1Substitute
the signal generator for the function generator.
2Substitute the function generator for the signal generator.