TB 9-6625-2097-24
Test instrument
parameters
Performance specifications
Recorders, Ink Writing, Model 2200S (Series)
Input signal
Range: 0 to 5 V at dc to 50 Hz
Frequency response
Range: 100 div at 100 mm amplitude, dc to 30 Hz
1 div
Accuracy:
Range: 50 div at 50 mm amplitude, dc to 50 Hz
1 div
Accuracy:
Overshoot
Range: 100 mm amplitude
Accuracy:
Less than 8 ms
Range: 50 mm amplitude
Accuracy:
Less than 5 ms overshoot
less than 1% between 10 to 90% FS
100, 115, 200, and 230 V, 10%, 50 to 400 Hz
Input power
RMS Converter, Model 13-4618-10
Converter output
Range: 0 to 5 V dc linear
0.2% of reading when reference to input
Accuracy:
voltage with accuracy of 0.05%
Converter input
Range: X1 V 0.05 to 10 V
X100 V 5 to 1000 V
Frequency Deviation Converter, Model 13-4618-00
Range: 0.5, 1, 2.5, and 5 Hz at 50 or 60 Hz center frequency
Frequency Deviation
5, 10, 25, and 50 Hz at 400 Hz center frequency
Output voltage:
2.5 V dc FS proportional to frequency
deviation at each range switch setting
Accuracy:
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286, AN/GMS-287, or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or exceed
the minimum use specifications listed in table 2. The accuracies listed in table 2 provided a four-to-
one ratio between the standard and TI. Where the four-to-one ratio cannot be met, the actual
accuracy of the equipment is shown in parenthesis.
5. Accessories Required. The accessories required for this calibration are common usage
accessories, issued as indicated in paragraph 4 above, and are not listed in the calibration
procedure. The following peculiar accessories are also required for this calibration:
Recorder, Gould, Model 2200S (Series) and Extender Card and Assembly, Gould, P/N
887291(supplied with TI).