Figure 1. Model 2200S (Series) - top view.
SECTION IV. CALIBRATION PROCESS FOR RMS CONVERTER, MODEL 13-4618-10

Calibration Procedure For Recorders, Ink Writing, Model 2200S (Series: With Rms Converter, Model 13-4618-10; And Frequency Deviation Converter, Model 13-4618-00 Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
img
. . . . . .
TB 9-6625-2097-24
Figure 2. Model 2200S (Series) - rear view.
b. Adjustments
(1) Set calibrator output for 2.500 V dc.
(2) Adjust CH 1 R201 SPAN (CH. 2 R201 SPAN) (fig. 1) for a 25 mm pen deflection
to the left on TI.
9. Overshoot
a. Performance Check
(1) Adjust function/arbitrary generator for 1 Hz square wave at 4 V p-p.
(2) Connect function/arbitrary generator to SIGNAL INPUT 1 (fig. 2).
(3) Turn VARIABLE SPEED control to fully cw.
If waveform on TI does not
indicate less than 1 percent overshoot, perform b below.
(4) Repeat technique of (2) above for SIGNAL INPUT 2 (fig. 2).
6