SECTION IV. CALIBRATION PROCESS FOR RMS CONVERTER, MODEL 13-4618-10
Converter Output

Calibration Procedure For Recorders, Ink Writing, Model 2200S (Series: With Rms Converter, Model 13-4618-10; And Frequency Deviation Converter, Model 13-4618-00 Page Navigation
  3    4    5    6    7  8  9    10    11    12  
img
. . . .
TB 9-6625-2097-24
13. Equipment Setup
a. Remove protective cover from TI.
b. Slide all manual ink valves with tubing attached to ON position on recorder.
c. Connect TI to recorder, using extender.
d. Set front panel controls as listed in (1) through (7) below:
(1) FULL SCALE switch to 1.
(2) ZERO SUPPRESSION switch OFF.
(3) ZERO SUPPRESSION control ccw to 0.
(4) FILTER switch OFF.
(5) MODE switch to X1 VOLTS.
(6) SENSITIVITY control fully cw.
(7) RANGE switch to X1.
e. Set DC INPUT switch S201 to left (fig. 3).
f. Energize equipment and allow 15 minutes for warm-up and stabilization.
Figure 3. Model 13-4618-10 - right side view.