TB 9-6625-2114-35
5. Accessories Required. The
accessories required for the
calibration are
common
usage accessories, issued as indicated in paragraph 4 above and are not listed in this
calibration procedure. The following peculiar accessory is also required for this
calibration: Probe, T-connector, P/N 11042/455A (8899531).
Table 2. Minimum Specifications of
Equipment Required
Minimum use
Manufacturer and model
Common name
specifications
(part number)
CALIBRATOR
Range: 0 to 315 V ac at 400 Hz
John Fluke, Model 5720A (p/o MIS-
891 to 909 V ac at 10 kHz
35947); w/power amplifier, John
Accuracy: 0.75%
Fluke, Model 5725A (5725A)
Range: 0.3 V ac; 20 Hz to 1 MHz
Accuracy: 0.5%
Range: 0 to 1030 V dc
Accuracy: 0.5%
Range: 0 to 150 MA dc
Accuracy: 0.75%
MEASURING RECEIVER
Hewlett-Packard, Model 8902A
Amplitude: 0.3 V
(8902A) and Sensor Module, Model
1.0% 60 to 90 MHz
Accuracy:
11722A (11722A)
4.5% 150 to 700 MHz
Range: 4.9 to 6.4 V ac
John Fluke, Model 8840A/AF-05
0 to 51.5 mV dc
(AN/GSM-64D)
Accuracy: 0.75%
(SG-1207/U)
Amplitude: 0.3 V
Range: 0 to 1.1 M Ω
Biddle-Grey, Model 71-631
Accuracy: 1.25%
(7910328)
Range: 0 to 10 M Ω
Beckman, Model CR10M (8598965)
Accuracy: 1.25%
MODEL 410B
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration process. Personnel should become familiar with the applicable sections
before beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by