TB 9-6625-2114-35
Calibrator indications
Test instrument
(V dc)
RANGE
switch settings
Indications
Min
Max
RX1 1V
.6
.570
.630
RX1 1V
.2
.170
.230
RX10 3V
3.0
2.910
3.090
RX100 10V
1.01
9.70
10.30
RX1K 30V
3.0
29.10
30.90
RX10K 100V
1.0
97.0
103.0
RX100K 300V
3.0
291.0
309.0
RX1M 1000V
dc only
1.0
970.0
1030.0
1After
this check, and only if dc voltage divider is supplied with TI, set calibrator to 9.0 V. Record TI
indication for use in (9) below.
NOTE
Perform (6) through (9) below only if dc voltage divider is
supplied with TI.
(6) Set RANGE switch to RX100 10V.
(7) Connect dc voltage divider to DC probe. Connect dc voltage divider and
COMMON lead together and adjust ZERO ADJ for 0 meter indication.
(8) Connect dc voltage divider and COMMON lead to calibrator.
(9) Adjust calibrator for TI indication recorded in table 3.
Calibrator will
indicate between 855.0 and 945.0 V.
b. Adjustments
(1) Set calibrator to 1.000 V.
(2) Adjust R32 CAL DC (fig. 1) for a 1 V indication on TI (R).
6