• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Swept Frequency and Marker Accuracy
Swept Frequency and Marker Accuracy - Continued

Calibration Procedure For Sweep Oscillator, Modles 8350A, 8350B, And Plug-In Models 8325A, 83592A, And 83592A Opt 002 Page Navigation
  23    24    25    26    27  28  29    30    31    32    33  
img
.
TB 9-6625-2124-24
Table 6. Marker Accuracy
Microwave measurement system
Test
indication
Instrument
(GHz)
Marker
Min
Max
M2
3.850
4.150
M3
7.850
8.150
M4
13.850
14.150
M5
16.850
17.150
(9) Adjust synthesized signal generator frequency control to align the birdie with
the end of TI sweep as indicated on oscilloscope display.
(10) Synthesized signal generator will indicate between 17.450 and 17.550 GHz.
(11) Press function keys and enter corresponding data on keyboard as listed in as
listed in (a) through (d) below:
START key, DATA ENTRY 10 MHz.
(a)
STOP key, DATA ENTRY 2.4 GHz.
(b)
M1 key, DATA ENTRY 1 GHz.
(c)
M2 key, DATA ENTRY 2 GHz.
(d)
(12) Repeat (6) above.
(13) Repeat technique of (7) above.
Synthesized signal generator will indicate
between 0.974 and 1.026 GHz.
(14) Repeat technique of (6) and (7) above for M2 marker.
Synthesized signal
generator will indicate between 1.974 and 2.026 GHz.
(15) Repeat (9) above.
(16) Synthesized signal generator will indicate between 2.385 and 2.415 GHz.
(17) Press function keys and enter corresponding data on keyboard as listed in as
listed in (a) through (j) below:
START key, DATA ENTRY 2.4 GHz.
(a)
STOP key, DATA ENTRY 7 GHz.
(b)
M1 key, DATA ENTRY 3 GHz.
(c)
M2 key, DATA ENTRY 6 GHz.
(d)
M3 key.
(e)
OFF key.
(f)
M4 key.
(g)
OFF key.
(h)
M5 key.
(i)
OFF key.
(j)
(18) Repeat (6) above.
(19) Repeat technique of (7) above.
Synthesized signal generator will indicate
between 2.957 and 3.043 GHz.
(20) Repeat technique of (6) and (7) above for M2 marker.
Synthesized signal
generator will indicate between 5.957 and 6.043 GHz.
28


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.