Table 6. Marker Accuracy
Figure 9. Switch A3S1 location.

Calibration Procedure For Sweep Oscillator, Modles 8350A, 8350B, And Plug-In Models 8325A, 83592A, And 83592A Opt 002 Page Navigation
  24    25    26    27    28  29  30    31    32    33    34  
img
. .
TB 9-6625-2124-24
(21) Adjust synthesized signal generator to align the birdie with the start of TI sweep
as indicated on oscilloscope display.
(22) Synthesized signal generator will indicate between 2.380 and 2.420 GHz.
(23) Repeat (9) above.
(24) Synthesized signal generator will indicate between 6.980 and 7.020 GHz.
(25) Press function keys and enter corresponding data on keyboard as listed in (a)
through (d) below:
START key, DATA ENTRY 7 GHz.
(a)
STOP key, DATA ENTRY 13.5 GHz.
(b)
M1 key, DATA ENTRY 8 GHz.
(c)
M2 key, DATA ENTRY 12 GHz.
(d)
(26) Repeat (6) above.
(27) Repeat technique of (7) above.
Synthesized signal generator will indicate
between 7.9425 and 8.0575 GHz.
(28) Repeat technique of (6) and (7) above for M2 marker.
Synthesized signal
generator will indicate between 11.9425 and 12.0575 GHz.
(29) Repeat (21) above.
(30) Synthesized signal generator will indicate between 6.975 and 7.025 GHz.
(31) Repeat (9) above.
(32) Synthesized signal generator will indicate between 13.475 and 13.525 GHz.
(33) Press function keys and enter corresponding data on keyboard as listed in (a)
through (d) below:
START key, DATA ENTRY 13.5 GHz.
(a)
STOP key, DATA ENTRY 17.5 GHz.
(b)
M1 key, DATA ENTRY 15 GHz.
(c)
M2 key, DATA ENTRY 17 GHz.
(d)
(34) Repeat (6) above.
(35) Repeat technique of (7) above.
Synthesized signal generator will indicate
between 14.9375 and 15.0625 GHz.
(36) Repeat technique of (6) and (7) above for M2 marker.
Synthesized signal
generator will indicate between 16.9375 and 17.0625 GHz.
(37) Repeat (21) above.
(38) Synthesized signal generator will indicate between 13.470 and 13.530 GHz.
(39) Repeat (9) above.
(40) Synthesized signal generator will indicate between 17.470 and 17.530 GHz.
b. Adjustments
(1) Connect multimeter to A6TP9 (+) and A6TP10 (-) (fig. 7).
(2) Set switch A3S1 (fig. 9) position 1 to the OPEN (UP) position.