TB 9-6625-2134-24
a. Performance Check
(1) Connect oscilloscope calibrator to RF INPUT and to TI EXT TRIG input (rear
panel of TI).
(2) Position controls as listed in (a) through (h) below:
(a) FREQUENCY SPAN/DIV control to 100 MHz.
(b) FREQUENCY control to 1 MHz.
(e) REFERENCE LEVEL control to -20 dBm.
(f) TIME/DIV control to 50 ms.
(g) TRIGGERING pushbutton to EXT.
(h) MIN RF ATTEN control to 20 dB.
control is reduced to zero.
(4) Adjust horizontal position control to aline a marker on first graticule line.
Check displacement of markers from their respective positions over center 8 divisions.
Individual marker displacement must not exceed 2 minor divisions (5 percent).
(5) Repeat technique of (3) and (4) above, using settings listed in table 7. Individual
marker displacement must not exceed 2 minor divisions (5 percent).
Table 7. Time/Division Accuracy Check.
Test instrument
Oscilloscope calibrator
TIME/DIV switch settings
marker setting
(ms)
(ms)
20
20
10
10
5
5
2
2
(6) Deactivate DIGITAL STORAGE pushbuttons.
control for a suitable display of time markers.
(8) Repeat technique of (4) above for TIME/DIV ranges of 1 ms to 20 s, applying
appropriate oscilloscope calibrator time marker output. Individual marker displacement
must not exceed 2 minor divisions.
(9) Disconnect oscilloscope calibrator and connect equipment as shown in figure 30.
(10) Set oscilloscope calibrator to .1 s markers.
(11) Position controls as listed in (a) through (e) below: