Calibration Procedure For Oscilloscope An/Usm-488 And Type 2235 Page Navigation
  1  2  3    4    5    6    7  
img
. . . . . . . .
TB 9-6625-2139-24
Table 1. Calibration Description - Continued
Test instrument
Performance specifications
parameters
Horizontal
A sweep timing
Range: 0.5 s/div to 0.05 Ps/div
Accuracy: 2%
Range: (X10 mag): 50 ms/div to 5 ns/div
Accuracy: 3%
B sweep timing
Range: 50 ms/div to 0.05 Ps/div
Accuracy: 2%
Range: (X10 mag): 5 ms/div to 5 ns/div
Accuracy: 3%
Sweep linearity
Accuracy: 5% (measured over any 2 of the center
8 divisions)
Deflection (X-Axis)
Range: 2 mV/div to 5 V/div
Accuracy: 3%
A trigger sensitivity
Frequency
10
MHz
60  MHz
100  MHz
1
Internal
1.0 div
1.5 div
0.35 div
150  mV2
External
35
mV
120
mV
B trigger sensitivity
Internal only
0.35 div
1.0 div
1.5 div
Calibrator amplitude
Range: 0.5 V
Accuracy: 2%3
1
0.3 division for type 2235.
2
200 mV for type 2235.
3
+5% for type 2235.
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure.  This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286; AN/GSM-287; or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI.  Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required.  The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessory is also required for this calibration:
standardizer, 5-80 pF.