TB 9-6625-2139-24
(3) Press oscilloscope calibrator VOLTAGE pushbutton to illuminate green LED.
(4) Adjust A TRIGGER LEVEL and POSITION controls, as necessary, to view
(5) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton to
obtain 5 divisions of vertical display. Oscilloscope calibrator err display will indicate
(6) Repeat technique of (4) and (5) above for settings listed in table 3. Oscilloscope
calibrator err display will indicate within limits specified in table 3; if not, perform
Oscilloscope
Oscilloscope
Test instrument
Test instrument
calibrator Err
calibrator
Test instrument
divisions of
VOLTS/DIV
display indications
VOLTAGE
adjustments
vertical deflection
switch settings
output settings
(r %)
b(1) thru (41)
2
m
10
mV
5
2
5
m
20
mV
4
2
10
m
50
mV
5
2
20
m
0.1
V
5
2
50
m
0.2
V
4
2
0.1
0.5
V
5
2
0.2
1
V
5
2
0.5
2
V
4
2
1
5
V
5
2
2
10
V
5
2
5
20
V
4
2
(7) Set VERTICAL MODE CH 1 BOTH CH 2 switch to CH 2 and move TI
connections at CH1 to CH2.
(8) Ensure CH 2 VOLTS/DIV switch is set to 2m.
(10) Adjust A TRIGGER LEVEL and POSITION controls, as necessary, to
(11) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton to
obtain 5 divisions of vertical display. Oscilloscope calibrator err display will indicate
(12) Repeat technique of (10) and (11) above for settings listed in table 4.
Oscilloscope calibrator err display will indicate within limits specified in table 4; if not,
6