SECTION II. EQUIPMENT REQUIREMENTS
Vertical

Calibration Procedure For Oscilloscope An/Usm-488 And Type 2235 Page Navigation
  1    2  3  4    5    6    7    8  
img
. . . . . . . . .
TB 9-6625-2139-24
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
OSCILLOSCOPE CALIBRATOR
Volts out:
Fluke, Model 5820A-5C-GHZ
Range: 10 mV to 20 V
(5820A-5C-GHZ)
Accuracy: 0.5%
Time markers:
Range: 5 ns/D to 0.5 s/D
Accuracy: 0.5%
Sine wave frequency:
Range: 50 kHz to >100 MHz
MULTIMETER
Range: -8.64 to < 0.1 V dc
Fluke, Model 8840A/AF05
Accuracy: 0.12%
(AN/GSM-64D)
SECTION III
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure.  Additional
maintenance information is contained in the manufacturer's manuals for this TI.
d. When indications specified in paragraphs 8 through 11 are not within tolerance,
perform the power supply check prior to making adjustments. After adjustments are made,
repeat paragraphs 8 through 11.  Do not perform power supply check if all other
parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to TI.
7. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance check where applicable.
a. Remove protective cover from TI only when necessary to make adjustments.
Replace cover after completing the adjustments.