TB 9-6625-2183-24
Table 1. Calibration Description - Continued
Test instrument
Performance specifications
parameters
Range: 1 s/div to 0.5 s/div
Horizontal
timing
18 steps in a 1-2-5 sequence, X10 mag extends
fastest sweep to 0.1 s/div
Accuracy: Unmag: 3% over center 8 div except
disregard first 0.5 s of sweep
Mag: X10 5% over center 8 div except exclude
first 10 div and all past 90 div
Gain
Range: 5 mV/div to 10 V/div 11 steps in a 1-2-5 sequence
Accuracy: 4%
Within 0.1 div or less bowing or tilt of a displayed
vertical or horizontal line
Range: 2, 20, 200, 1000 V
Accuracy: 0.1%, of reading 2 counts
Range: 2, 20, and 700 V ac, 40 to 500 Hz
Accuracy: 5%, 10 counts
A
Range: 2, 20, 200, 2000 k
ccuracy: 0.6%, 3 counts
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Sets AN/GSM-286, AN/GSM-287 or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessory is required for this calibration: DC
power supply; Elgar, model DCS40-30EM10.