SECTION II. EQUIPMENT REQUIREMENTS
Equipment Setup

Calibration Procedure For Oscilloscope Type, 305 Dmm Page Navigation
  1    2  3  4    5    6    7    8  
img
. . . . . .
TB 9-6625-2183-24
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
AUTOTRANSFORMER
Range: 105 to 125 V ac
Ridge, Model 9020A
Accuracy:  1%
(9020A)
Fluke, Model 5720A (5720A) (p/o MIS-
CALIBRATOR
Range: 1.9 to 700 V ac
Accuracy:  0.125%
35947)
Range: 1.896 to 1003 Vdc
Accuracy:  0.0525%
MULTIMETER
Range: -3 to +150 V dc
Agilent, Model 3458A (3458A)
Accuracy:  0.175%
Range: 0.1 to 50 s
OSCILLOSCOPE
Agilent, OS-303/G (OS-303/G)
OSCILLOSCOPE
Marker:
Fluke, Model 5820A-5C-GHZ
Range: 1 s/div to 0.5
CALIBRATOR
(5820A-5C-GHZ)
s/div
Accuracy:  0.75%
Voltage:
Range: 0.03 to 60 V
Accuracy:  0.75%
Edge:
Range: 1, 0.1, 0.01 ms
Risetime:  17 ns
Flatness:  0.75%
Level Sine:
Range: 0.25 to 5 MHz 0.05
MHz reference
Flatness:  3%
A
RESISTANCE
Biddle-Gray, Model 71-631
Range: 1.9, 19, 190, 900 k
ccuracy:  0.125%
STANDARD NO. 1
(7910328)
A
RESISTANCE
Beckman, Model CR10M (8598965)
Range: 1000 k
ccuracy:  0.125%
STANDARD NO. 2
SECTION III
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
c. Unless otherwise specified, verify the results of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure.  Additional
maintenance information is contained in the manufacturer's manual for this TI.