RF Output Frequency Accuracy
Table 13. RF Output Frequency Accuracy

Calibration Procedure For Radar Test Set An/Tpm-25A (Ts-3394A) Page Navigation
  20    21    22    23    24  25  26    27    28    29    30  
img
. .
TB 9-6625-2211-24
(8) Adjust signal generator No. 1 output frequency to 1085 MHz and adjust output
level controls and TI SUM ATTEN control until a birdie is present on oscilloscope Channel 2.
Figure 8. RF output frequency - equipment setup.
Figure 9. RF output timing markers - oscilloscope display.
25