TB 9-6625-2213-24
(7) Adjust A4R5 (fig. 4) for a TI meter indication of 0 dB.
Figure 4. Test instrument - adjustments locations.
(8) Replace channel B probe with channel A probe in equipment setup.
(9) Repeat (1) through (8) above until both indications are within tolerance.
NOTE
If (1) through (9) above cannot be obtained, adjust A11R24 (fig.
4) for peak AMPLITUDE meter indication and perform the IF
section adjustments as listed in TM 11-6625-2856-14 or the
manufacturer's manual.
13