Output Distortion - Continued
Table 6. Sine Wave Amplitude Accuracy 50 Hz to 99.99 kHz

Calibration Procedure For Function Generator Sg-1288/G Page Navigation
  10    11    12    13    14  15  16    17    18    19    20  
img
. . . . .
TB 9-6625-2233-24
Figure 4. A7 phase lock loop circuit card adjustment location.
13. Sine Wave Amplitude
a. Performance Check
(1) Connect UNBALANCED output to multimeter input using a 50 Ω
feedthrough termination.
(2) Press pushbuttons as listed in (a) through (h) below:
(a) CALIBRATE
(allow
enough
time
to
complete
and
verify
AUTOCALIBRATED is displayed).
(b) RESET.
(c) AMPLITUDE.
(d) 1. 1 1.
(e) ENTER.
(f) FREQUENCY.
(g) 50.
(h) ENTER.
(3) Multimeter will indicate between minimum and maximum values listed in table 6.
(4) Repeat technique of (2) (c) through (h) and (3) above substituting settings and
indications listed in table 6.