Table 7. Sine Wave Amplitude Accuracy 100 kHz to 4.999 MHz
Table 9. Triangle Wave Amplitude Accuracy 50 Hz to 99.99 kHz

Calibration Procedure For Function Generator Sg-1288/G Page Navigation
  13    14    15    16    17  18  19    20    21    22    23  
img
. . . .
TB 9-6625-2233-24
Table 8. Sine Wave Amplitude Accuracy 5 MHz to 19.99 MHz
Oscilloscope indications
Test instrument
(V p-p)
Output
Frequency
(V p-p)
Min
Max
5
MHz
.099
.086
.112
5
MHz
.2
.174
.226
5
MHz
.999
.877
1.12
5
MHz
2
1.75
2.25
5
MHz
15
13.2
16.8
15.99  MHz
.099
.086
.112
15.99  MHz
.2
.174
.226
15.9
MHz
.999
.877
1.12
15.99  MHz
2
1.75
2.25
15.99  MHz
15
13.2
16.8
19.99  MHz
.099
.068
.130
19.99  MHz
.2
.138
.262
19.99  MHz
.999
.697
1.30
19.99  MHz
2
1.39
2.61
19.99  MHz
15
10.5
19.5
b. Adjustments. No adjustments can be made.
14. Triangle Wave Amplitude
a. Performance Check
(1) Connect UNBALANCED output to multimeter input using 50 Ω feedthrough termination.
(2) Press pushbuttons as listed in (a) through (i) below:
(a) CALIBRATE
(allow
enough
time
to
complete
and
verify
AUTOCALIBRATED is displayed).
(b) RESET.
(c) FUNCTION (triangle wave).
(d) AMPLITUDE.
(e) .009.
(f) ENTER.
(g) FREQUENCY.
(h) 50.
(i) ENTER.
(3) Multimeter will indicate between 1.94 and 3.25 mV rms.
(4) Repeat technique of (2) (c) through (i), and (3) above substituting settings and
indications listed in table 9.