Table 6. Sine Wave Amplitude Accuracy 50 Hz to 99.99 kHz
Triangle Wave Amplitude

Calibration Procedure For Function Generator Sg-1288/G Page Navigation
  12    13    14    15    16  17  18    19    20    21    22  
img
. . .
TB 9-6625-2233-24
Table 7. Sine Wave Amplitude Accuracy 100 kHz to 4.999 MHz
Measuring receiver indications
Test instrument
(mV)
Output
Frequency
(V p-p)
Min
Max
100
kHz
.099
33.2
36.8
100
kHz
.2
67.17
74.24
100
kHz
.999
338
368
100
kHz
2
675
739
100
kHz
15
5086
5518
999.9
kHz
.099
33.2
36.8
999.9
kHz
.2
67.17
74.24
999.9
kHz
.999
338
368
999.9
kHz
2
675
739
999.9
kHz
15
5086
5518
1
MHz
.099
32.9
37.1
1
MHz
.2
66.46
74.94
1
MHz
.999
335
372
1
MHz
2
668
746
1
MHz
15
5034
5571
4.999  MHz
.099
32.9
37.1
4.999  MHz
.2
66.46
74.94
4.999  MHz
.999
335
372
4.999  MHz
2
668
746
4.999  MHz
15
5034
5571
(12) Reduce TI output to minimum.  Disconnect UNBALANCED output from
measuring receiver sensor module and connect UNBALANCED output to oscilloscope
Channel 1 input using a 50 Ω feedthrough termination.
(13) Press pushbuttons as listed in (a) through (h) below:
(a)
FREQUENCY.
(b)
5.
(c)
EXP.
(d)
6.
(e)
ENTER.
(f)
AMPLITUDE.
(g)
.099.
(h)
ENTER.
(14) Set oscilloscope Channel 1 inputs for DC Coupling, 1M  Input, scaling for
approximately 4 or 5 divisions displayed, sweep speed for approximately 4 or 5 cycles
displayed and volts pk-pk measurement.
(15) Oscilloscope will indicate within minimum maximum limits as listed in table 8.
(16) Repeat technique of (13) through (15) above substituting settings and
indications listed in table 8.
18