TB 9-6625-2233-24
Table 9. Triangle Wave Amplitude Accuracy 50 Hz to 99.99 kHz
Test instrument
Output
(Hz)
(V p-p)
Min
Max
50
.01
1.93
mV
3.84
mV
50
.099
26.9
mV
30.3
mV
50
.1
26.8
mV
30.9
mV
50
.999
.279
V
.298 V
50
1
.277
V
.300 V
50
9.99
2.79
V
2.97
V
50
10
2.79
V
2.98
V
50
15
4.19
V
4.47
V
99.99 k
.01
1.93
mV
3.84
mV
99.99 k
.099
26.9
mV
30.3
mV
99.99 k
.1
26.8
mV
30.9
mV
99.99 k
.999
.279
V
.298 V
99.99 k
1
.277
V
.300 V
99.99 k
9.99
2.79
V
2.97
V
99.99 k
10
2.79
V
2.98
V
99.99 k
15
4.19
V
4.47
V
(5) Reduce TI output to minimum. Set the measuring receiver to measure power.
(6) Zero and calibrate measuring receiver sensor module.
(7) Set the measuring receiver to measure power in millivolts.
(8) Disconnect UNBALANCED output from multimeter and connect UNBALANCED
output to measuring receiver sensor module (do not use the 50 Ω feedthrough termination).
(9) Press pushbuttons as listed in (a) through (h) below:
(a)
(b)
100.
(c)
EXP.
(d)
3.
(e)
ENTER.
(f)
AMPLITUDE.
(g)
.099.
(h)
ENTER.
(10) Measuring receiver power meter will indicate within minimum maximum limits
20