TB 9-6625-2233-24
Table 12. Square Wave Amplitude Accuracy 50 Hz to 99.99 kHz - Continued
Test instrument
Frequency
Output
rms indications
(Hz)
(V p-p)
Min
Max
99.9 k
.099
46.5
mV
52.5
mV
99.9 k
.1
46.5
mV
53.5
mV
99.9 k
.999
.483 V
.516 V
99.9 k
1
.480 V
.520 V
99.9 k
9.99
4.84
V
5.15
V
99.9 k
10
4.84
V
5.16
V
99.9 k
15
7.27
V
7.74
V
(5) Reduce TI output to minimum.
(6) Zero and calibrate measuring receiver sensor module.
(7) Set the measuring receiver to measure power in millivolts.
(8) Disconnect UNBALANCED output from multimeter and connect UNBALANCED
output to measuring receiver sensor module (do not use the 50 Ω feedthrough termination).
(9) Press pushbuttons as listed in (a) through (h) below:
(b) 100.
(c) EXP.
(d) 3.
(e) ENTER.
(f) AMPLITUDE.
(g) .099.
(h) ENTER.
(10) Measuring receiver will indicate within minimum maximum limits as listed in
(11) Repeat technique of (9) (a) through (h) and (10) above substituting settings and
Table 13. Square Wave Amplitude 100 kHz to 4.999 MHz
Measuring receiver indications
Output
(mV)
(Hz)
(V p-p)
Min
Max
100
k
.099
45.52
53.48
100
k
.1
45.50
54.50
100
k
.999
472.5
526.5
100
k
1
470.0
530.0
999.9
k
.099
45.52
53.48
999.9
k
.1
45.50
54.50
999.9
k
.999
472.5
526.5
999.9
k
1
470.0
530.0
1
M
.099
45.03
53.97
1
M
.1
45.00
55.00
4.999 M
.099
45.03
53.97
4.999 M
.1
45.00
55.00
23