Table 9. Triangle Wave Amplitude Accuracy 50 Hz to 99.99 kHz
Square Wave Amplitude

Calibration Procedure For Function Generator Sg-1288/G Page Navigation
  15    16    17    18    19  20  21    22    23    24    25  
img
. . . .
TB 9-6625-2233-24
Table 10. Triangle Wave Amplitude Accuracy 100 kHz to 4.999 MHz
Test instrument
Measuring receiver indications
(mV)
Frequency
Output
(Hz)
(V p-p)
Min
Max
100
k
.099
26.28
30.87
100
k
.1
26.27
31.47
100
k
.999
272.8
304.0
100
k
1
271.4
306.0
999.9
k
.099
26.28
30.87
999.9
k
.1
26.27
31.47
999.9
k
.999
272.8
304.0
999.9
k
1
271.4
306.0
1
M
.099
26.00
31.16
1
M
.1
25.98
31.75
1
M
.999
270.0
306.8
1
M
1
268.5
308.9
4.999  M
.099
26.00
31.16
4.999  M
.1
25.98
31.75
4.999  M
.999
270.0
306.8
4.999  M
1
268.5
308.9
(11) Repeat technique of (9) (a) through (h) and (10) above substituting settings and
indications listed in table 10.
(12) Reduce TI output to minimum.  Disconnect UNBALANCED output from
measuring receiver sensor module and connect UNBALANCED output to oscilloscope
Channel 1 input using a 50 Ω feedthrough termination.
(13) Press pushbuttons as listed in (a) through (h) below:
(a)
FREQUENCY.
(b)
1.
(c)
EXP.
(d)
6.
(e)
ENTER.
(f)
AMPLITUDE.
(g)
9.99.
(h)
ENTER.
(14) Set oscilloscope Channel 1 inputs for DC Coupling, 1M  Input, scaling for
approximately 4 or 5 divisions displayed, sweep speed for approximately 4 or 5 cycles
displayed and volts pk-pk measurement.
(15) Oscilloscope will indicate within minimum maximum limits as listed in table 11.
(16) Repeat technique of (13) through (15) above substituting settings and
indications listed in table 11.
21