Table 13. Square Wave Amplitude 100 kHz to 4.999 MHz
Dc Offset and Attenuator

Calibration Procedure For Function Generator Sg-1288/G Page Navigation
  18    19    20    21    22  23  24    25  
img
. . .
TB 9-6625-2233-24
(12) Reduce TI output to minimum.  Disconnect UNBALANCED output from
measuring receiver sensor module and connect UNBALANCED output to oscilloscope
Channel 1 input using a 50 Ω feedthrough termination.
(13) Press pushbuttons as listed in (a) through (h) below:
(a)
FREQUENCY.
(b)
1.
(c)
EXP.
(d)
6.
(e)
ENTER.
(f)
AMPLITUDE.
(g)
.999.
(h)
ENTER.
(14) Set oscilloscope Channel 1 inputs for DC Coupling, 1M  Input, scaling for
approximately 4 or 5 divisions displayed, sweep speed for approximately 4 or 5 cycles
displayed and volts pk-pk measurement.
(15) Oscilloscope will indicate within minimum maximum limits as listed in table 14.
(16) Repeat technique of (13) through (15) above substituting settings and
indications listed in table 14.
Table 14. Square Wave Amplitude 1 to 19.99 MHz
Oscilloscope indications
Frequency
Output
(V p-p)
(Hz)
(V p-p)
Min
Max
1
M
.999
.935  mV
1.06
V
1
M
1
.930  mV
1.07
V
1
M
9.99
9.38
V
10.6 V
1
M
10
9.38
V
10.6
V
1
M
15
14.1
V
15.9
V
4.999 M
.999
.935  mV
1.06
V
4.999 M
1
.930  mV
1.07
V
4.999 M
9.99
9.38
V
10.6
V
4.999 M
10
9.38
V
10.6
V
4.999 M
15
14.1
V
15.9
V
5
M
.009
5.92
mV
12.1
mV
5
M
.01
5.8
mV
14.2
mV
5
M
.099
84
mV
114
mV
5
M
.1
84
mV
116
mV
5
M
.999
875
mV
1.12
V
5
M
1
870
mV
1.13
V
5
M
9.99
8.78
V
11.2
V
24