Figure 6. Automatic amplitude discrimination test - equipment setup.

Calibration Procedure For Microwave Frequency Counter, Model 5352B Option 001 Page Navigation
  9    10    11    12    13  14
img
. . . . .
TB 9-6625-2247-40
13. Power Supply
NOTE
Do not perform the power supply check if all other parameters
are within tolerance.
a. Performance Check. Connect negative lead of multimeter to TI chassis ground
and positive lead of multimeter to +3 VOLT TEST POINT (fig. 1). If multimeter does not
indicate between 2.98 and 3.02 V dc, perform b below.
b. Adjustments. Adjust A3R4 (fig.1) for a multimeter indication between 2.98 and
3.02 V dc. (R)
14. Final Procedure
a. Deenergize and disconnect all equipment
b. Annotate and affix DA label in accordance with TB 750-25.
15/(16 blank)