Timebase Stability
Figure 2. Input 2 sensitivity (50 ?) - equipment setup.(5) Insert a 50 ? feed throug

Calibration Procedure For Microwave Frequency Counter, Model 5352B Option 001 Page Navigation
  1    2    3    4  5  6    7    8    9    10  
img
. . . .
TB 9-6625-2247-40
Figure 1. Test instrument - adjustment locations.
b. Adjustments. No further adjustments can be made.
9. Input 2 Sensitivity Test (10 Hz to 525 MHz)
a. Performance Test
(1) Press 50Ω key on FUNCTION/DATA keyboard.
(2) Connect equipment as shown in figure 2.
(3) Adjust synthesized signal generator frequency to 50 MHz and RF output
amplitude controls for a +1.0 dBm indication on power meter. If TI does not indicate a
repeatable count 1 Hz of synthesized signal generator frequency, perform b below.
NOTE
In technique (3) above, the power meter indication combined
with the 20 dB attenuator (fixed) and power splitter test report
values must equal -19 dBm.
(4) Repeat technique (3) above for synthesized signal generator frequencies of 100,
200, 400, and 525 MHz.