(3) Adjust synthesized signal generator frequency to 12,370 MHz and RF output
level controls for a -16.5 dBm indication on power meter.
on with equal brightness. (R)
(5) Press RESET/LOCAL key.
(6) Adjust synthesized signal generator RF level output controls for a -17 dBm
indication on power meter. YELLOW LED (fig. 1) will be on and GREEN LED (fig. 1) will
be off.
(7) Adjust synthesized signal generator RF level output controls for a -16 dBm
indication on power meter. YELLOW LED (fig. 1) will be off, and GREEN LED (fig. 1) will
be on.
NOTE
If techniques (6) and (7) above fail,
repeat techniques (1) through (5) above.
(8) Press the following keys as listed in (a) through (e) below. TI will display an IF
frequency of 70 MHz.
SET/ENTER.
(a)
DIAGNOSTICS.
(b)
6.
(c)
3.
(d)
SET/ENTER.
(e)
(9) Connect synthesized signal generator RF output through 20 dB attenuator
(fixed) to power meter.
(10) Adjust synthesized signal generator frequency to 209 MHz and RF output
amplitude controls for a -20 dBm indication on power meter.
(11) Disconnect synthesized signal generator RF output and 20 dB attenuator (fixed)
from power meter.
(12) Connect synthesized signal generator RF output through 20 dB attenuator
(fixed) to INPUT 1 and connect power meter to IF OUT on TI rear panel.
(13) Press SET/ENTER key.
(14) Press the following keys on FUNCTION/DATA keyboard listed in (a) through
(c) below:
(a) DIAGNOSTICS.
(b) 9.
(c) 7.
(15) Press SET/ENTER key. TI display will indicate ADJUST IF 175 FILTER D97.
(16) Adjust A6C7 (fig. 1) for a minimum indication on power meter. (R)
(17) Set POWER switch to STBY, and disconnect power meter from IF OUT (rear panel).
9