TB 9-6625-2259-24
(4) Press F1 M3 key and set F1 M3 display to 0.010 GHz by pressing 10 with
DATA ENTRY keys and then by pressing MHz/dB/mS DATA ENTRY key.
(5) Press F2 M6 key and set F2 M6 display to 2.000 GHz by pressing 2 with
DATA ENTRY key and then by pressing GHz/dBm/SEC DATA ENTRY key.
(6) Press SWEEP TIME key and set dB SWEEP display to 1.0 second by pressing
1 with DATA ENTRY key and then by pressing GHz/dBm/SEC DATA ENTRY key.
(7) Press LEVEL key and set LEVEL display to +13.0 dBm by pressing 13 with
DATA ENTRY keys and then by pressing GHz/dBm/SEC DATA ENTRY key.
(8) Slowly rotate INCR-DECR control cw until UNLEVELED indicator blinks on,
or until LEVEL display indicates +15.0 dBm.
NOTE
If LEVEL display indication for (8) above is +15.0 dBm,
proceed to (10) below.
(9) Slowly rotate INCR-DECR control ccw until UNLEVELED indicator quits
blinking. Wait for 3 sweeps to complete without the UNLEVELED indicator blinking
before proceeding to (10) below.
(10) Press TRIGGER MANUAL SWEEP key and rotate TRIGGER MANUAL
SWEEP control fully ccw.
(11) Configure measuring receiver to measure frequency and allow measuring
receiver to acquire carrier frequency.
(12) Connect multimeter (dc mode) to V/GHz OUTPUT (rear panel).
(13) Adjust trigger manual sweep control for a multimeter indication as close as
possible to value recorded in 7 h (8) above.
(14) Configure measuring receiver to measure RF power.
(15) Slowly rotate TRIGGER MANUAL SWEEP control cw until multimeter
indicates same value as recorded in 7 h (10) above. If measuring receiver RF power display
lowest level indications are not equal to or greater than +13 dBm, perform b below.
(16) Slowly rotate TRIGGER MANUAL SWEEP control cw until multimeter
indicates same value as recorded in 7 h (11) above. If measuring receiver RF power display
indications are not equal to or greater than +13 dBm, perform b below.
(17) Press FREQUENCY RANGE F1-F2 key and then press TRIGGER AUTO key.
(18) Press F1 M3 key and set F1 M3 display to 2.000 GHz by pressing 2 with
DATA ENTRY keys and then by pressing GHz/dBm/SEC DATA ENTRY key.
(19) Press F2 M6 keys and set F2 M6 display to 18.000 GHz by pressing 18 with
DATA ENTRY keys and then by pressing GHz/dBm/SEC DATA ENTRY key.
(20) Press SWEEP TIME key and set SWEEP display to 1.0 second by pressing 1
with DATA ENTRY key and then by pressing GHz/dBm/SEC DATA ENTRY key.
17