(15) Press F1 M3 key and set F1 M3 display to 2.000 GHz by pressing 2 with DATA
ENTRY key and then by pressing GHz/dBm/SEC DATA ENTRY key.
(16) Press F2 M6 key and set F2 M6 display to 18.000 GHz by pressing 18 with
DATA ENTRY keys and then by pressing GHz/dBm/SEC DATA ENTRY key.
(17) Set LEVEL display to +10.0 dBm by pressing 10 with DATA ENTRY keys and
then by pressing GHz/dBm/SEC DATA ENTRY key.
(18) Adjust TRIGGER MANUAL SWEEP CONTROL for a multimeter indication
(19) Adjust INCR-DECR control for a measuring receiver FREQUENCY/LEVEL
indication of +10.0 dBm.
(20) Configure measuring receiver to measure frequency at 2 GHz and allow
measuring receiver to acquire carrier frequency.
(21) Configure measuring receiver to measure RF power.
Measuring receiver
indication will be between +8.9 and +11.1 dBm.
(22) Repeat technique of (18), (20), and (21) above for remaining multimeter
indications and measuring receiver CARRIER FREQ settings listed in table 9.
Measuring receiver
frequency settings
(V dc)
(GHz)
_________1
2
3.0
3
4.0
4
5.0
5
6.0
6
7.0
7
8.0
8
9.0
9
10.0
10
11.0
11
12.0
12
13.0
13
14.0
14
15.0
15
16.0
16
17.0
17
________2
18
1Adjust
TRIGGER MANUAL SWEEP control for a multimeter
indication equal to value recorded in 7 h (11) above.
2Adjust TRIGGER MANUAL SWEEP control for a multimeter
indication equal to value recorded in 7 h (12) above.
(23) Record 10 dB attenuation (fixed) value from test report at 10 MHz and 2 GHz.