• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Output Power Variations Test
Output Power Variations Test - Continued

Calibration Procedure For Sweep Generator Sg-1206( )/U Page Navigation
  15    16    17    18    19  20  21    22    23    24    25  
img
. . .
TB 9-6625-2259-24
(15) Press F1 M3 key and set F1 M3 display to 2.000 GHz by pressing 2 with DATA
ENTRY key and then by pressing GHz/dBm/SEC DATA ENTRY key.
(16) Press F2 M6 key and set F2 M6 display to 18.000 GHz by pressing 18 with
DATA ENTRY keys and then by pressing GHz/dBm/SEC DATA ENTRY key.
(17) Set LEVEL display to +10.0 dBm by pressing 10 with DATA ENTRY keys and
then by pressing GHz/dBm/SEC DATA ENTRY key.
(18) Adjust TRIGGER MANUAL SWEEP CONTROL for a multimeter indication
equal to value listed in table 9.
(19) Adjust INCR-DECR control for a measuring receiver FREQUENCY/LEVEL
indication of +10.0 dBm.
(20) Configure measuring receiver to measure frequency at 2 GHz and allow
measuring receiver to acquire carrier frequency.
(21) Configure measuring receiver to measure RF power.
Measuring receiver
indication will be between +8.9 and +11.1 dBm.
(22) Repeat technique of (18), (20), and (21) above for remaining multimeter
indications and measuring receiver CARRIER FREQ settings listed in table 9.
Table 9. Output Power Variations Test
Measuring receiver
Multimeter indications
frequency settings
(V dc)
(GHz)
_________1
2
3.0
3
4.0
4
5.0
5
6.0
6
7.0
7
8.0
8
9.0
9
10.0
10
11.0
11
12.0
12
13.0
13
14.0
14
15.0
15
16.0
16
17.0
17
________2
18
1Adjust
TRIGGER MANUAL SWEEP control for a multimeter
indication equal to value recorded in 7 h (11) above.
2Adjust TRIGGER MANUAL SWEEP control for a multimeter
indication equal to value recorded in 7 h (12) above.
(23) Record 10 dB attenuation (fixed) value from test report at 10 MHz and 2 GHz.
(24) Press FREQUENCY RANGE F1 F2 key.
21