Flatness Test - Continued
Output Attenuator Test - Continued

Calibration Procedure For Sweep Generator Sg-1206( )/U Page Navigation
  20    21    22    23    24  25  26    27  
img
. . . .
TB 9-6625-2259-24
Table 10. Flatness Test
Measuring receiver
Multimeter indications
frequency settings
(V dc)
(GHz)
________1
2
3.0
3
4.0
4
5.0
5
6.0
6
7.0
7
8.0
8
9.0
9
10.0
10
11.0
11
12.0
12
13.0
13
14.0
14
15.0
15
16.0
16
17.0
17
_________2
18
1Adjust
TRIGGER MANUAL SWEEP control for a multimeter
indication equal to value recorded in 7 h (11) above.
2Adjust TRIGGER MANUAL SWEEP control for a multimeter
indication equal to value recorded in 7 h (12) above.
b. Adjustments. No adjustments can be made.
16. Output Attenuator Test
a. Performance Check
(1) Press RESET key.
(2) Press LEVEL key and set LEVEL display to +13.0 dBm by pressing 13 with
DATA ENTRY keys and then by pressing GHZ/dBm/SEC DATA ENTRY key.
(3) Press FREQUENCY RANGE CW F1 key and set F1 M3 display to 1.000 GHz
by pressing 1 with DATA ENTRY key and then by pressing GHz/dBm/SEC DATA
ENTRY key.
(4) Zero and calibrate measuring receiver power sensor.
(5) Connect power sensor to TI RF OUTPUT 50 .
(6) Configure measuring receiver to measure frequency at and allow measuring
receiver to acquire carrier frequency.
(7) Configure measuring receiver to measure RF power. Measuring receiver
indication will be within limits specified in first row of table 11.
26