Output Attenuator Test - Continued

Calibration Procedure For Sweep Generator Sg-1206( )/U Page Navigation
  22    23    24    25    26  27
img
. . . . . . . . . .
TB 9-6625-2259-24
(16) Repeat technique of (12) above for remaining TI LEVEL settings listed in table
13. Measuring receiver indication will be within limits specified in table 13.
Table 13. Output Attenuator Test @ 18 GHz
Measuring receiver
Test instrument
indication
LEVEL settings
(dB)
(dBm)
Min
Max
10
7.4
12.6
3
1.4
5.6
(17) Press LEVEL key and adjust INCR-DECR control for 0.0 dBm indication on
measuring receiver.
(18) Using standard tuned level measurement techniques, verify the measuring
receiver indicates within limits specified in first row of table 14.
(19) Repeat technique of (9) and (10) above for remaining TI LEVEL settings listed
in table 14. Measuring receiver will indicate within limits specified in table 14.
Table 14. Output Attenuator Test @ 18 GHz
Measuring receiver
Test instrument
indication
LEVEL settings
(dB)
(dBm)
Min
Max
0
-2.6
2.6
-10
-12.6
-7.4
-20
-22.6
-17.4
-30
-32.6
-27.4
-40
-42.6
-37.4
-50
-52.6
-47.4
-60
-62.6
-57.4
b. Adjustments. No adjustments can be made.
17. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.