TB 9-6625-2259-24
(5) Press F2 M6 key and set F2 M6 display to 2.000 GHz by pressing 2 with DATA
ENTRY key and then by pressing GHz/dBm/SEC DATA ENTRY key.
(6) Press LEVEL key and set LEVEL display to 0.0 dBm by pressing 0 with
DATA ENTRY keys and then by pressing GHz/dBm/SEC DATA ENTRY key.
(7) Press TRIGGER MANUAL SWEEP key and rotate TRIGGER MANUAL
SWEEP control fully ccw.
(8) Connect measuring receiver power sensor to RF OUTPUT 50 .
(9) Configure measuring receiver to measure frequency at and allow measuring
receiver to acquire carrier frequency.
(10) Connect multimeter (dc mode) to V/GHz OUTPUT (rear panel).
(11) Adjust trigger manual sweep control for a multimeter indication as close as
possible to value recorded in 7 h (8) above.
(12) Configure measuring receiver to measure RF power.
(13) Adjust INCR-DECR control for a 0.0 dBm indication on measuring receiver.
(14) Slowly rotate TRIGGER MANUAL SWEEP control cw and until multimeter
indicates same value as recorded in 7 h (10) above. Measuring receiver will indicate
between -1.5 and +1.5 dBm.
(15) Slowly rotate TRIGGER MANUAL SWEEP control cw until multimeter
indicates same value as recorded in 7 h (11) above. Measuring receiver will indicate
between -1.5 and +1.5 dBm.
(17) Press F1 M3 key and set F1 M3 display to 2.000 GHz by pressing 2 with DATA
ENTRY keys and then by pressing GHz/dBm/SEC DATA ENTRY key.
(18) Press F2 M6 keys and set F2 M6 display to 18.000 GHz by pressing 18 with
DATA ENTRY keys and then by pressing GHz/dBm/SEC DATA ENTRY key.
(19) Adjust TRIGGER MANUAL SWEEP control for a multimeter indication equal
(20) Configure measuring receiver to measure frequency at and allow measuring
receiver to acquire carrier frequency.
(21) Configure measuring receiver to measure RF power.
Measuring receiver
indications will be between +1.5 and -1.5 dBm.
(22) Repeat technique of (20) and (21) for remaining multimeter indications and
measuring receiver CARRIER FREQ settings listed in table 10.
25