Maximum Output Power Test - Continued
Figure 4. A1A4A4 component location diagram.

Calibration Procedure For Sweep Generator Sg-1206( )/U Page Navigation
  12    13    14    15    16  17  18    19    20    21    22  
img
. . .
TB 9-6625-2259-24
(21) Press LEVEL key and set LEVEL display to +10.0 dBm by pressing 10 with
DATA ENTRY keys and then by pressing GHz/dBm/SEC DATA ENTRY key.
(22) Slowly rotate INCR-DECR control cw until UNLEVELED indicator blinks on.
(23) Slowly rotate INCR-DECR control ccw until UNLEVELED indicator quits
blinking. Wait for 3 sweeps to complete without the UNLEVELED indicator blinking
before proceeding to (24) below.
(24) Press TRIGGER MANUAL SWEEP key.
(25) Adjust TRIGGER MANUAL SWEEP control for a multimeter indication
equal to the value listed in table 8.
(26) Configure measuring receiver to measure frequency at 2 GHz.
(27) Configure measuring receiver to measure RF power. If measuring receiver RF
power indication is not equal to or greater than +10 dBm, perform b below.
(28) Repeat technique of (25) through (27) above for remaining multimeter
indications and measuring receiver frequency settings listed in table 8.
Table 8. Maximum Output Power Test
Measuring receiver
Multimeter indications
frequency settings
(GHz)
(V dc)
1
______
2
3.0
3
4.0
4
5.0
5
6.0
6
7.0
7
8.0
8
9.0
9
10.0
10
11.0
11
12.0
12
13.0
13
14.0
14
15.0
15
16.0
16
17.0
17
__________2
18
1Adjust
TRIGGER MANUAL SWEEP control for a multimeter
indication equal to value recorded in 7 h (11) above.
2Adjust TRIGGER MANUAL SWEEP control for a multimeter
indication equal to value recorded in 7 h (12) above.
b. Adjustments
(1) Press POWER pushbutton to OFF position and remove TI top cover. Remove TI
board cover screw holding CCA boards.
(2) Replace TI top cover and do not fasten down (this step allows for easy removal
and replacement of TI top cover).