TB 9-6625-2280-40
(14) Press RF detector B pushbutton to ON and RF detector A to OFF.
(15) Connect RF detector B to CALIBRATOR 1 GHz output.
(16) Repeat (5) through (11) above using RF detector B.
b.
Adjustments
NOTE
Perform first multimeter hookup and adjustments on A6 board
and when instructed by text in (4) below, perform multimeter
hookup and adjustments on A7 board.
(1) Connect multimeter (dc mode) LO lead to A6TP1 (A7TP1) (fig. 3).
(2) Connect multimeter (dc mode) HI lead to A6TP2 (A7TP2) (fig. 3).
(3) Adjust A6R13 (A7R13) (fig. 3) for a multimeter indication between 9.997 and
10.003 V dc.
(4) Repeat (1) through (3) above for A7 board.
(5) Disconnect multimeter leads from TI.
15